Fritsch PULVERISETTE 6 classic line Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Maszyny specjalne Fritsch PULVERISETTE 6 classic line. FRITSCH PULVERISETTE 6 classic line User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1

Instruction manual Gas-Pressure- and Temperature- Measuring-system „GTM II“ for Laboratory-Planetary Mill “pulverisette 5” and Laboratory-Mono-Mill

Strona 2

GTM-System page 7 2.4 Loading of the battery in the transmitter (accu in transmitter unit) 2.4.1 voltage drop warning • In case that the voltage

Strona 3

GTM-System page 8 2.4.3 Accumulator-lifespan • The lifespan of the accumulators decreases with overheating of the transmitter unit. • The accumu

Strona 4 - GTM-System page 1

GTM-System page 9 2.6 Software installation 2.6.1 Prerequisites for the installation of the soft-ware • Common PC with MS-Windows (95 or highe

Strona 5 - 1.3.2 Method of Operation

GTM-System page 10 2.7.1 Menu Options / configuration broadcast-unit • Serial number: Can only be changed when administrator mo-dus is being acti

Strona 6 - GTM-System page 3

GTM-System page 11 2.7.2 Setup – Configuration Set values in the configuration must be transmitted to the trans-mitter unit for activation. Connect

Strona 7 - 1.4 Technical data

GTM-System page 12 3 Execution of Measurement 3.1 Filters • Loosen the two slotted screws (under the transmitter unit). • Remove plate • Place

Strona 8 - GTM-System page 5

GTM-System page 13 Mass balance may be made by: volume of the grinding bowl (ml) material weight without grinding balls and sample (gr.) order n

Strona 9 - 2 Installation

GTM-System page 14 • With „CHECK IN“ you are being reminded to pull the plug out of the transmitter unit. Doing so will activate it and the data tr

Strona 10 - 2.4.2 Recharging

GTM-System page 15 3.2.2 Ending Measuring • Select from the menu Measurement the command Close or press the button close recording (arrow). • Swi

Strona 11 - 2.4.3 Accumulator-lifespan

GTM-System page 16 4 Measurement evaluation Measuring values and additional information, which has been saved in ASCII-format may be used for furthe

Strona 12 - 2.6 Software installation

Edition 08/2002 Index 004 checked and released Fritsch GmbH Laborgerätebau Industriestrasse 8 D - 55743 Idar-Oberstein Phone: +49 (0)6784/ 70-

Strona 13 - GTM-System page 10

GTM-System page 17 5 Detailed information 5.1 GTM Software The updating of the diagram during the measurement takes place within a second after arri

Strona 14 - 2.7.2 Setup – Configuration

GTM-System page 18 5.2.2 Stand by with configuration Stand by, extends the operation time of the transmitter unit with one ac-cumulator-charge. In

Strona 15 - 3 Execution of Measurement

GTM-System page 19 7 Maintenance Before working on the appliance disconnect from mains! The GTM system is maintenance-free, in general. Neverthele

Strona 16 - GTM-System page 13

GTM-System page 20 9 Troubleshooting Checklist Malfunction Possible cause Elimination of error Light POWER is off No connection to power Plug in

Strona 17 - GTM-System page 14

GTM-System page 21 11 A1 11.1 Installation manual Software GTM II Analysis software runs with WINDOWS 95/98/ME/NT40/2000© operating systems 11.1.

Strona 18 - 3.2.2 Ending Measuring

GTM-System page 22 5. Installation directory can be changed with Browse Æ next 6. 6. Program file may be changed Æ next 7. Start installati

Strona 19 - 4 Measurement evaluation

GTM-System page 23 8. Installation of files is running 9. Program file and icon 10. Finish!

Strona 20 - 5 Detailed information

Table of contents page 1 General Information / Introduction...1 1.1 Notes about the Operating Manual ...

Strona 21 - 6 Cleaning

GTM-System page 1 1 General Information / Introduction 1.1 Notes about the Operating Manual • The copyrights to these technical documents remain

Strona 22 - 8 Warranty

GTM-System page 2 1.3 Brief Description of the Instrument 1.3.1 Applications The Measuring System GTM grasps simultaneously gas pressure and tempe

Strona 23 - 9 Troubleshooting Checklist

GTM-System page 3 1.3.3 The measure principle The measure principle is based on it that pressure and temperature are directly proportional in an i

Strona 24 - 11.1.1 Installation steps:

GTM-System page 4 1.4 Technical data The evaluation software runs with WINDOWS 95/98/ME/NT40/2000© operating systems Dimensions receiver 1100

Strona 25 - GTM-System page 22

GTM-System page 5 1.5 General Safety Instructions • The instrument can only to be used for the purpose described • Use only original accessorie

Strona 26 - GTM-System page 23

GTM-System page 6 1.6 Electrical Safety Please pay attention to the input voltage of the receivers! Please plug the cables only into the correct so